72610 DO Sensor
3-2610-31 to 9900 Set Up
1. Wire the 9900, 2610-31, and 8058 as shown in the diagrams.
2. On the 9900 press and
hold the ENTER key for 2
seconds.
3. Press the key to select the
INPUT menu item.
4. Press the ENTER key to
access the INPUT menus.
5. Press the key to select the
TYPE menu item.
6. Press the key to change
the TYPE selection.
7. Enter the password code.
8. Repeatedly press the key
until the “4-20 mA INPUT”
menu item is ashing and
then press the ENTER key.
9. The 9900 will prompt “All
settings will be reset”, press the
key to select “Yes” and then press the ENTER key.
10. The 9900 will return to the View Mode display.
11. Press and hold the ENTER key for 2 seconds.
12. Press the key to select the INPUT menu item.
13. In the NAME menu item, press the key to change the
displayed name from “4-20 mA INPUT” to a more descriptive
name (e.g., "DISSOLVED O2") and press ENTER when done.
14. Press the key to select SENSOR UNIT menu item.
15. Press the to change the label from UNIT to MG/L or PPM
and press ENTER.
16. Press the key and ensure the 4 mA VALUE is set to 0.0000.
17. Press the key and change the 20 mA VALUE from
5.0000 to 20.000 and press ENTER.
18. Press both the and keys simultaneously to return to
the Menu.
19. Press the to select the LOOP menu and press ENTER.
20. Set the 4 MA SET POINT to your desired value. The 2610 is
factory set for a 0 to 20 mg/L output. Press ENTER when done.
21. Press the key to select the 20 MA SET POINT and set to your desired value. The 2610 is factory set for a
0 to 20 mg/L output. Press ENTER when done.
22. Press both the and keys simultaneously to return to the Menu.
23. Press the key twice to select the OPTION menu and press ENTER.
24. Press the key twice to select the SET BAR MIN option. Change this option if desired. The 2610 is factory
set for a 0 to 20 mg/L output. Press the ENTER key when done.
25. Press the key to select the SET BAR MAX option. Change this option if desired. The 2610 is factory set for
a 0 to 20 mg/L output. Press ENTER when done.
26. Press both the and keys simultaneously to return to the Menu.
27. ENTER the other menus and set the unit as desired for your application.
28. Press both the and keys simultaneously to return to the View Menu.
RED
DO Sensor
3-2610-31
24 VDC
Regulated
Power Supply
BLACK
Input
4-20 mA
Output
S
3
L
SHIELD
(no connection)
Signet 8058 i-Go™
4-20 mA to S
3
L Converter
+GF+
4-20 mA input S
3
L Output
V–
V+
BROWN
GREEN
BLUE
WHITE
No connection
8058-1
PWR–
RED
WHITE
SHIELD
BLACK
DATA
GND
SHLD
V+
9900 S
3
L Inputs9900 Power Inputs
LOOP–
LOOP+
PWR+
PWR–
LOOP–
LOOP+
PWR+
DATA
GND
V+
SHLD
+
-
3 2 1
6 5 4
7 8 9
Loop1
PWR
N/C
S L
DATA
3
Loop2
PWR
+GF+ SIGNET
BROWN
24 VDC
Regulated
Power Supply
DO
Sensor
1
3-2610-31
No connection
9900 Power Inputs
9900 S
3
L Inputs
8058-2
BLACK (5 VDC)
BLACK
BLACK
RED
RED
WHITE (GND)
RED (S
3
L)
Technical Notes:
The cable length from the 8058 to the 9900 must not exceed 60 m (200 ft).
When using the 8058-2 with the 9900, connect the loop source to 8058-2 Channel 1 input
ONLY as shown in the gure.
See the 8058 manual for more information.