52610 DO Sensor
Wiring and Set Up
3-2610-31 to 8900 Set Up
1. Connect the 8058, 2610-31, and 8900 as shown in the diagram.
2. Press and hold ENTER key for 5 seconds.
3. The display should ash “System Setup”.
4. Press ENTER key.
5. Scroll to the desired channel for Dissolved
Oxygen.
6. Press the key.
7. Enter the password.
8. Repeatedly press the key until the lower
line of the display reads “Other (4-20)”.
9. Press ENTER key.
10. The 8900 will warn that “Channel Data will
be Reset. Are you sure?” Press the key
so that “Yes” is ashing and press ENTER
key.
11. Simultaneously press the and keys to
return to the Menu Directory.
12. Press the key to select the “Channel
Settings” menu and press ENTER key to
select.
13. Use the key to select the channel used
in step 5.
14. Press the key to change the label and press ENTER when done.
15. Press the key to select the
abbreviation.
16. Press the key to change the
label and press ENTER when
done. Press the key to select
the Units.
17. Press the key to change
the Units. The 2610 shipped
with default units of mg/L or
ppm, either can be used. Press
ENTER when completed.
18. Press the key to select the 4
mA Set Point.
19. Press the key to change the set point. By default the 4 mA set point for the 2610 is 0.0. Press the ENTER
key when complete.
20. Press the key to select the 20 mA Set Point.
21. Press the key to change the set point. By default the 20 mA set point for the 2610 is 20.0. Press the
ENTER key when complete.
22. Press the key to select the Decimal location.
23. Press the key and change the decimal position if desired. Press ENTER when complete.
24. Simultaneously press the and keys to exit out of the channel menu.
25. Continue programming other options in the 8900 or simultaneously press the and keys to return to the
View Mode.
6
7
8
9
10
11
12
13
14
+
-
3 2 1
6 5 4
7 8 9
Loop1
PWR
N/C
S L
DATA
3
Loop2
PWR
+GF+ SIGNET
8058-2
BLACK (5 VDC)
BLACK
BLACK
BLACK
RED
RED
WHITE (GND)
RED (S
3
L)
24 VDC
Regulated
Power Supply
DO
Sensor
1
3-2610-31
No connection
DO
Sensor
2
3-2610-31
No connection
8900 Sensor Inputs
BROWN
BROWN
RED
DO Sensor
3-2610-31
24 VDC
Regulated
Power Supply
BLACK
Input
4-20 mA
Output
S
3
L
SHIELD
(no connection)
Signet 8058 i-Go™
4-20 mA to S
3
L Converter
+GF+
4-20 mA input S
3
L Output
V–
V+
BROWN
GREEN
BLUE
WHITE
No connection
8058-1
8900 Sensor Inputs
6
7
8
9
10
11
12
13
14
BLACK
RED
WHITE
SHIELD
(no connection)
Wiring and Set Up