6
2610 DO Sensor
1. Connect the 3-2610-41, and 8900 as shown in the diagram
NOTE: The wiring of the 3-2610-41 is non-standard:
• Red wire is connected to 12 to 24 VDC
• White wire is connected to S
3
L Data
• Black wire is connected to VDC Ground
• A jumper wire must be connected between the VDC Ground and S
3
L Gnd.
2. Press and hold Enter key for 5 seconds. The display will read “System Setup”.
3. Press Enter key.
4. Scroll to the channel that will be dissolved oxygen.
5. Press the key.
6. Enter the password.
7. Repeatedly press the key until the lower
line of the display reads “Other (4-20)”.
8. Press Enter key.
9. The 8900 will warn that “Channel Data will
be Reset. Are you sure?” press the key so
that “Yes” is ashing and press Enter key.
10. Simultaneously press the and keys to
return to the Menu Directory.
11. Press the key to select the menu “Channel
Settings” and press Enter key to select.
12. Use the key to select the channel used in
step 4.
13. Press the key to change the label and
press Enter when done.
14. Press the key to select the abbreviation.
15. Press the key to change the label and
press Enter when done.
16. Press the key to select the Units.
17. Press the key to change the Units. The
2610 shipped with default units of mg/L, or
ppm, either can be used. Press Enter when
completed.
18. Press the key to select the 4mA Set Point.
19. Press the key to change the set point. By default the 4mA set point for the 2610 is 0.0. Press the Enter key
when complete.
20. Press the key to select the 20mA Set Point.
21. Press the key to change the set point. By default the 20 mA set point for the 2610 is 20.0. Press the Enter
key when complete.
22. Press the key to select the Decimal location..
23. Press the key and change the decimal position if desired. Press Enter when complete.
24. Simultaneously press the and keys to exit out of the channel menu.
25. Continue programming other options in the 8900 or simultaneously press the and keys to return to
View Mode.
Wiring and Set Up
3-2610-41 to 8900 Set Up
The 3-2610-41 Optical Dissolved Oxygen Sensor with S
3
L was designed to emulate a 4 to 20 mA current
input device, e.g., Signet 3-8058, on the 8900 controller. This allows the 3-2610-41 to be backward
compatible with all existing 8900 controllers.
I/O Module 3-8900.401-X
1
2
3
4
5
6
7
8
9
10
11
12
13
14
+5VDC (Black)
Freq. Input (Red)
GND (Shield)
+5VDC (Black)
Freq. Input 2 (Red)
S L (Red)
GND (White/Shield)
+5VDC (Black)
S L (Red)
GND (White/Shield)
3
3
Analog Output 1
Analog Output 2
(if applicable)
(if applicable)
Frequency
Input
1
Frequency
Input 2
OR
S
3
L
Input
2
S
3
L
Input
1
+
-
+
-
OUT 2 OUT 1
POWER
COMM PORT /
OUT 4 OUT 3
RELAY 2 RELAY 1
SENSOR INPUTS
C
NO
NC
NO
NC
C
C
NO
NC
NO
NC
C
N
L
+
-
DC ONLY
+
-
+
-
24 VDC
Regulated
Power Supply
V+
V-
No Connection:
Green
Blue
Brown
3-8900
3-2610-41
3-2610-41
Channel 1
Channel 2
White
Red
Red
Black
Black
Vedere la pagina 5