Signet 2610 Process Optical Dissolved Oxygen Sensor
3-2610.090 Rev. C 08/13
*3-2610.090*
English
Operator's Manual
Your new RDO
®
Pro is a rugged, reliable sensor designed to
deliver accurate dissolved oxygen (DO) data across a wide
measuring range while reducing maintenance costs. It features
the latest optical technology for DO measurement.
Introduction
The RDO Pro system consists of the following:
• 10 m (32.8 ft.) cable with stripped and tinned ends
• Black sensor body with removable nose cone
• Optical DO sensing cap
• Titanium thermistor
Sensor cap
(not visible)
Cable end, stripped and tinned
Nose cone
Thermistor
The
he
T
rmi
rmi
mi
sto
sto
r
System Description
Dimensions
Serial Numbers
The instrument part number, code number and serial number is
engraved on the side of the unit.
RDO is a registered trademark of In-Situ
®
Inc., Fort Collins, CO USA
20.3 cm
(8.0 in.)
68.6 mm
(2.7 in.)
43.7 mm
(1.72 in.)
The RDO Pro system is available in two versions. The 3-2610-31
has 4 to 20mA current loop output, Modbus RS485, and SDI-12
connectivity. The 3-2610-41 includes S
3
L communications to
support both the 8900 and 9900 Signet products in place of the
SDI-12 output.
Vedere la pagina 0
1 2 3 4 5 6 ... 11 12

Sommario

Pagina 1 - *3-2610.090*

Signet 2610 Process Optical Dissolved Oxygen Sensor3-2610.090 Rev. C 08/13*3-2610.090*EnglishOperator's ManualYour new RDO® Pro is a rugged

Pagina 2 - Calibration

102610 DO SensorNotes

Pagina 3 - Care and Maintenance

112610 DO SensorSpecifi cationsGeneralSensor type: ...Luminescent dissolved oxygen sensorTransmitter/local display: .

Pagina 4 - Sensor Cap Replacement

Ordering Information2610 Optical DO SensorMfr. Part No Code Description3-2610-31 159 001 753 Optical DO Sensor (0-20 ppm) with Modbus, SDI, an

Pagina 5 - Wiring and Set Up

22610 DO SensorIntroductionUnpacking the Sensor1. Remove the RDO® sensor from the box and other packaging materials.2. Unscrew the nose cone from t

Pagina 6 - 3-2610-41 to 8900 Set Up

32610 DO SensorSensor DeploymentSensor DeploymentThe cable end of the RDO® Pro is internally threaded (1¼ - 11½ NPT) and can be attached to a external

Pagina 7 - 3-2610-31 to 9900 Set Up

42610 DO SensorCare and MaintenanceSensor Cap ReplacementReplace the Sensor CapThe sensor cap has a 1-year life after the instrument takes its fi rst r

Pagina 8 - 3-2610-41 to 9900 Set Up

52610 DO SensorWiring and Set Up3-2610-31 to 8900 Set Up1. Connect the 8058, 2610-31, and 8900 as shown in the diagram.2. Press and hold ENTER key f

Pagina 9 - 3-2610 to Modbus Set Up

62610 DO Sensor1. Connect the 3-2610-41, and 8900 as shown in the diagramNOTE: The wiring of the 3-2610-41 is non-standard:• Red wire is connected t

Pagina 10 - 2610 DO Sensor

72610 DO Sensor3-2610-31 to 9900 Set Up1. Wire the 9900, 2610-31, and 8058 as shown in the diagrams.2. On the 9900 press and hold the ENTER key for

Pagina 11 - Specifi cations

82610 DO Sensor1. Wire the 9900 and 3-2610-41 as shown in the diagram.NOTE: The wiring of the 3-2610-41 is nonstandard:• 3-2610-41 Red wire is conne

Pagina 12 - Ordering Information

92610 DO SensorWiring and Set Up3-2610 to 4 to 20 mA Set Up3-2610 to Modbus Set UpRedBlackBlueGreenBrownWhite+ModbusRS485–ModbusRS485 3-2610-31or 3-26

Commenti su questo manuale

Nessun commento